PET project: Was able to incorporate timing information for all frames. For this non-gated acquisition the following logic was used:
(Frame Reference Time
(0x00541300)+Actual Frame Duration (0x00181242))/2
Now Dynamic PET datasets can be loaded, ROIs can be constructed and corresponding TACs (time activity curves) can be obtained.